Efficacy of Plasma Activation on Bleaching
Fatma Yilmaz1, Esra Uzer Celik2*, Utku Kursat Ercan3, Fatma Ibis4
Copyright : © 2018 Authors. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Objectives: This in vitro study aimed to evaluate the bleaching efficacy and color stability of nonthermal atmospheric pressure plasma (NAPP)-activated 40% hydrogen peroxide (HP) and deionized water (DW). Methods: Two hundred-forty bovine enamel−dentin samples were stained with a coffee solution before bleaching. Samples were randomly divided into 12 (n = 20): (1) HP for 5 minute (HP-5), (2) HP-10, (3) HP-20, (4) Plasma-HP-5, (5) P-HP-10, (6) P-HP-5/HP-15, (7) P-HP-10/HP-10, (8) P-DW-5, (9) P-DW-10, (10) DW-5, (11) DW-10, and (12) DW-20. After bleaching, the samples were re-stained for 48 h using red wine. The color differences measured using a dental spectrophotometer after bleaching and after re-staining were reported as ΔE1 and ΔE2, respectively. The data were analyzed using Kruskal−Wallis and Mann−Whitney U tests (p< 0.05).
Results: All plasma-activated groups revealed higher ΔE1 values than the control groups and similar or higher ΔE1 values than the nonactivated HP groups (p0.05). The HP-5 and HP-10 groups showed lower ΔE2 value than the other groups (p< 0.05), and ΔE2 value of the HP-20 group was similar to the other plasma-activated groups.
Conclusion: Nonthermal atmospheric pressure plasma treatment increased the bleaching efficacy of HP and gave similar bleaching efficacy of HP to DW with similar color stability.
1. Introduction
2. Materials And Methods
Group 1 (HP-5): 40% HP gel (Opalescence® Boost PF 40%, Ultradent Inc., South Jordan, UT, USA) was applied for 5 min according to the manufacturer’s instructions. A 0.5−1.0 mm-thick layer of gel was applied to the labial surfaces. Samples with HP on them were held in an incubator for 5 min during application at 37 °C.
Group 2 (HP-10): The same procedures as those used in group 1 were applied but for 10 min.
Group 3 (HP-20): The same procedures as those used in group 1 were applied but for 20 min.
Group 4 (P-HP-5): 40% HP was applied to the samples, and they were then treated for 5 min with nonthermal atmospheric DBD plasma at the previously described parameters.
Group 5 (P-HP-10): In this group, the same procedure in group 4 was performed but for 10 min.
Group 6 (P-HP-5/HP-15): After 40% HP was applied to the samples, it was activated by plasma performed as describe above for 5 min. Following the 5 min activation by plasma, the gel was not removed from the outer surface of the samples. Then, the samples were kept in an incubator at 37° C for 15 min.
Group 7 (P-HP-10/HP-10): Plasma was performed as describe above for 10 min. Following the 10 min activation by plasma, the gel was not removed from the outer surface of the samples. Then, the samples were kept in an incubator at 37° C for 10 min.
Group 8 (P-DW-5): DW was applied to the samples and then activated by plasma as describe above for 5 min. However, to ensure the electric field, the plasma source was switched off once every 1 min and then 1 drop of DW was placed onto the buccal surfaces of samples.
Group 9 (P-DW-10): In this group, the same procedure in group 8 was performed but for 10 min.
Group 10 (control group) (DW-5): The samples were kept in a DW-filled container for 5 min in an incubator at 37 °C as a control group of HP-5, P-HP-5, and P-DW-5.
Group 11 (control group) (DW-10): The samples were kept in a DW-filled container for 10 min in an incubator at 37 °C as a control group of the HP-10, P-HP-10, and P-DW-10 groups.
Group 12 (control group) (DW-20): The samples were kept in a DW-filled container for 20 min in an incubator at 37 °C as a control group of the HP-20, P-HP-5/HP-15, and P-HP-10/HP-10 groups.
After each application, the samples were cleaned with a brush under water for 30 s and evaluated. Following the completion of the bleaching process, the samples were polished using polishing discs (Sof-LexTM Polishing Discs, 3M ESPE, St. Paul, MN, USA) under water cooling. Then, fluoride gel (Elmex Gel, Gaba, Lörrach, Germany) was applied to the samples for 4 min.
The samples were kept in artificial saliva for 14 days after bleaching had been completed and subjected to red wine (Kaplankaya Wines, Sirince, Izmir, Turkey) for 48 h [20]. Following the re-staining process, the samples were cleaned by a toothbrush (Oral-B Indicator Plus 35, Gillette do Brasil Ltda., Manaus, Amazonas, Brazil) with a nonabrasive toothpaste (Colgate Total 12, Colgate-Palmolive Ltda., Osasco, São Paulo, Brazil). External discoloration was removed by pumice and polishing rubber cups.
Vita colors and L*a*b values of the samples were measured by a dental spectrophotometer (Spectroshade, MHT OpticReasearch AG, Niederhasli, Switzerland) after discoloration, bleaching, and re-staining. Measurements were performed in a dark room and repeated three times, but the average value was used for further analysis. The color changes occurring after bleaching (ΔE1) and after re-staining (ΔE2) were calculated by the following equation: ΔE1 = ((L2-L1)2+(a2-a1)2+(b2-b1)2)1/ 2 L1, a1, and b1: L, a, and b values after staining, L2, a2, and b2: L, a, and b values after bleaching; ΔE2 = ((L3-L2)2+(a3-a2)2+(b3-b2)2)1/2(L3, a3, and b3: L, a, and b values after re-staining, L2, a2, and b2: L, a, and b values after bleaching).
Three human central incisors were used to measure tooth temperature in each plasma-activated group. The temperature increase was measured by a fiber optic temperature measurement system (FTI-10 fiber optic signal conditioner, FOT-L-SD fiber optic temperature sensor; FISO Technologies Inc., Québec, Canada). The measurements were performed immediately after the bleaching process. The fiber optic temperature sensor was placed in contact with the buccal surface, and the tooth was properly fixed. The measurements were performed on each tooth at room temperature (25 °C).
Statistical analysis was conducted using the SPSS software system version 20.0 (IBM Corporation, New York, USA). Confidence analysis was performed by the Kolmogorov −Smirnov test (df=40), which proved that the values were not normally distributed. The Kruskal−Wallis test was used to make multiple comparisons of the groups. Binary comparisons within the group were performed by the Mann−Whitney U test. The results were evaluated at the 95% confidence interval and at the p < 0.05 level.
3. Results And Discussion
Color difference values of the tested bleaching methods and their statistical comparison results are given in Table 1 and 2.
Color improvement after bleaching (ΔE1 value) in the experimental groups was significantly higher than that in the control groups (p < 0.05). No statistical differences were observed among ΔE1 values of the control groups. In the experimental groups, the HP-5 and HP-10 groups revealed lower ΔE1 values than the other groups (p < 0.05). No differences were found in the ΔE1 values of the HP-20 and P-HP-5 groups, and the HP-20 group showed a lower ΔE1 value than that of the other plasma-activated groups (p < 0.05).
In the plasma-activated groups, the ΔE1 values of the P-HP-5 and P-HP-10 groups were lower than those of the P-DW-10 and P-HP-10/HP-10 groups (p < 0.05). No significant differences were observed in the other pairwise comparisons.
Re-staining after bleaching (ΔE2 values) in all of the experimental groups were significantly higher than those of the control groups except for the HP-5 group (p < 0.05). No statistical differences were observed in the control groups in terms of ΔE2 values. In the experimental groups, the HP-5 and HP-10 groups showed lower ΔE2 than the other groups (p < 0.05). The ΔE2 value of the HP-20 group was similar to those of the plasma activated groups.
In the plasma activated groups, the P-HP-10/HP-10 group showed statistically less re-staining than the P-HP-5, P-HP-10, P-DW-5, and P-DW-10 groups (p < 0.05). The ΔE2 value of the P-DW-5 group was also statistically lower than that of the P-HP-5 group (p < 0.05).
In the plasma-activated groups, temperature values before (t1) and after (t2) applications are given in Figure 3. The temperature did not increase to the intraoral temperature of 35° C−36° C in any of the groups [21].
According to the findings, the null hypothesis that “no difference would be found in bleaching efficacy and color stability between plasma-activated and nonactivated agents” was rejected. In the present study, NAPP increased the bleaching efficacy of HP and shortened the application time of the conventional in-office bleaching technique without light activation. The P-HP-5 group revealed 2.5 times more effective bleaching efficacy than HP-5. Similarly, the P-HP-10 group showed two times more effective bleaching than HP-10. We could not apply the plasma for 20 min because a long application time could cause excessive heat on the electrode surface that could lead to the corruption of the DBD electrode. However, we obtained similar bleaching efficacy in the HP-20 group when we applied 40% HP gel with plasma for 5 min (P-HP-5) and obtained better results than those of the HP-20 group when we applied 40% HP gel with the other plasma application modes (P-HP-10, P-HP-5/HP-15, and P-HP-10/HP-10).
In previous in vitro studies, NAPP was also used as a light source to increase the bleaching efficacy of HP and CP, as it is a nonthermal and nontoxic method [7,8, 22]. Park,et al. [23] reported that the NAPP jet enhanced the bleaching efficacy of 30% HP in intracoronal tooth bleaching. Nam et al. [24] observed that the NAPP jet-activated 16% CP was 3.7 times more effective than its nonactivated counterpart in tooth bleaching. They also compared the bleaching effectiveness of 16% CP activated by three different light sources (i.e., NAPP jet, PAC, and diode laser) and reported that the highest ∆E value was reached using the NAPP jet [25]. However, the plasma source used in these previous studies was different from the device used in the present study. We preferred the DBD air plasma over the NAPP jet as the jet system removes the gel away from the surface because of the high speed gas flow and the gel requires replacement at certain intervals [23,26]. Helium gas at a flow rate of 2 L min-1 is commonly used as feeding gas though the NAPP jet device. If helium is present even in just a small amount in the atmosphere, it changes the atmospheric conditions of the working environment. Moreover, inhaling helium can be dangerous because it is a simple asphyxiant and may displace oxygen needed for normal respiration [27].
Atomic oxygen radicals, which are highly reactive species, play an important role in the biomedical applications of plasma [28,29]. Plasma treatment provides bleaching by uncovering back the colorant organic molecules that are typically maintained on the enamel surface and by removing or denaturing proteins [13]. Furthermore, plasma may be effective in tooth bleaching by breaking the linked pairs of pigmented protein chains and modifying them to a single bond through the aid of OH ions released from plasma and quickly diffused to the enamel surface. Therefore, plasma creates a synergistic mechanism with bleaching agents and accelerates the bleaching process.
NAPP activation produced an effective tooth bleaching effect when it was applied with DW in the present study. The P-DW-5 group revealed 3.7 times more effective bleaching efficacy than the HP-5 group; similarly, the P-HP-10 group showed 3.1 times more effective bleaching than the HP-10 group. Furthermore, we revealed a better bleaching effect when we applied DW to all plasma activation modes (the P-DW-5, P-DW-10 groups) compared with the HP-20 group. In addition, the P-DW-10 group showed a superior bleaching effect with the P-HP-10/HP- 10 group to other plasma-activated groups. Similarly, tooth stain was removed within 10 min by a hybrid gas−water plasma in a previous study [9].
The effective bleaching effect of plasma-treated DW may be explained by the formation of reactive oxygen species, charged particles, and oxidizing radicals that are stronger than ozone [30, 31]. Both HP and OH radicals are generated after energetic electrons have interacted with the water molecules. OH radicals have been speculated to be mainly in charge of the tooth whitening process by the hybrid gas−liquid plasma [9]. In addition, plasma-activated DW has also been reported to possess an acidic feature [9, 32]. Shainsky et al. [33] showed that although the HP concentration in DW increased to 2000 mg/L at the beginning of the plasma application, it decreased to 10 mg/L after a 15 min application.
The oxidation ability of HP is high at low pH values [34]. Furthermore, superoxide anions occur in DW after plasma treatment. The oxidizing property of HP is enhanced by superoxide anions [33]. This information may explain the effective bleaching effect of plasma-activated DW.
It has been reported that enamel permeability increases in proportion to the concentration of bleaching gel [35]. The increased permeability of enamel may accelerate re-staining. In the present study, ΔE2 values of the experimental groups were statistically higher than those of the control groups except for the HP-5 group. This group showed similar ΔE2 values with the control groups probably because of its short application time. Similarly, bleached enamel by 35% HP re-stained more than nonbleached enamel [20, 36, 37]. Plasma activation of HP and DW for 5 min and 10 min revealed more re-staining than their nonactivated counterparts. The reason for this finding may be attributed to their effective bleaching effect. In previous studies, bleached teeth with high color difference values were also prone to re-staining [38, 39]. However, all plasma-activated groups produced similar re-staining with HP-20. This finding may indicate that plasma activation increases bleaching activity without any irreversible damage to the composition and surface morphology of enamel. In previous studies [36, 40], susceptibility to re-staining of enamel after bleaching depended on surface roughness, chemical composition, and dehydration amount occurring after the bleaching process.
In the present study, enamel surface morphology reached a maximum of 27°C after NAPP application. Similarly, the NAPP jet revealed no extra increase in temperature compared with the nonactivated bleaching groups or other light sources.[24-26] Sulieman et al. [41] claimed that bleaching gel and its water content act as a barrier preventing the increase in pulp temperature caused by light sources.
4. Conclusions